Report of laboratory and in-situ validation of micro-sensor for monitoring ambient air - Ozone micro-sensors, αSense, model B4 O3 sensors

Report of laboratory and in-situ validation of micro-sensor for monitoring ambient air - Ozone micro
Report of laboratory and in-situ validation of micro-sensor for monitoring ambient air - Ozone micro-sensors, αSense, model B4 O3 sensors
Abstract: The aim of this report was to evaluate and validate O3-B4 sensors of αSense within laboratory and field tests under ambient/indoor air conditions corresponding to a specific micro-environment: traffic station, urban/suburban areas. This report presents the evaluation of the performances and determination of the laboratory and field measurement uncertainty of the sensor values, compared to uncertainties fixed by the Data Quality Objective (DQO) of the European Air Quality Directive for indicative method. Further, procedures for the calibration of sensors able to ensure full traceability of measurements of sensors to SI units are developed.
Files
URI
Authors
Authors: 
SPINELLE Laurent, GERBOLES Michel, ALEIXANDRE Manuel
Publication Year
Publication Year: 
2014
Type

Type:

Science Areas
Keywords
Publisher
Publisher: 
Publications Office of the European Union
ISBN
ISBN: 
978-92-79-38538-4
ISSN
ISSN: 
1831-9424
DOI
Other Identifiers
Other Identifiers: 
EUR 26681