In Situ Gamma Spectroscopy of Spent Nuclear Fuel Using a CdTe Detector.

Abstract: 
Gamma spectroscopy using a room temperature CdTe detector has been carried out in the proximity of a spent nuclear fuel rod. In this context, a compact and transportable monitor, incorporating a planar effective shielding for the crystal against the intense background. Gamma spectra, with an energy resolution (FWHM) of 7.2 keV at 662 keV, were obtained in 600 s counting time, which would alow verification of the burnup to within 10% and cooling time of the fuel to within 3% of the declared values by applying specific isotopic correlations. The good spatial resolution has allowed the use of the monitor as a gamma scanning device.